Petroleum product analysis instrument
National consultation hotline: 0411-84794867

Ultraviolet fluorescence sulfur analyzer

DFYF-3000S

Introduction
summary:

SH / T 0689, ASTM d5453 light hydrocarbon and engine fuel and other oil determination of total sulfur content (UV fluorescence method). The instrument can be widely used in petroleum, chemical industry, electric power, coal, food, environmental protection and other fields.

Technical parameters:

L sample type: liquid, solid and gas (with corresponding injector)
USB port is used for data acquisition and transmission. The data acquisition card is imported, and the detection limit is 1uv.
L the key components of the instrument are: original imported components.
The operation interface of the instrument is infinitely extended without interruption, and the data can be searched infinitely. The spectrum can be called at any time.
High voltage range: DC500v ~ 700V, according to the measured concentration, set the required value through the operating system, the default value is about 550V.
L determination method: ultraviolet fluorescence (s)
L measurement range: 0.2 mg / L ~ 5000 mg / L ~ 100%
Detection limit: 0.2mg/l (s)
Temperature control range: room temperature ~ 1100 ℃, temperature control accuracy: 0.5% ± 3 ℃
L repeatability error: ① when 0.2mg/l ≤ sample concentration < 1.0, ≤± 0.1mg/l;
② When 1.0mg/l ≤ sample concentration < 100mg / L, CV ≤ 10%;
③ When 100mg / L ≤ sample concentration ≤ 10000mg / L, CV ≤ 5%.
L gas source requirements: argon: above 99.98%, oxygen: above 99.98%
L power supply: AC220 V ± 22 V, 50 Hz ± 0.5 Hz, power: ≤ 2 kW
L overall dimension: 680 × 480 × 400 (mm)
Label:

Service
Contact

Hotline

13304111506

Work Shift

from Monday to Friday

Office Number

0411-84794867

QR code