Petroleum product analysis instrument
National consultation hotline: 0411-84794867

DFYF-103Z Automated gasoline oxidation stability tester

DFYF-103Z

Introduction
Products Overviews:
 Automated gasoline oxidation stability tester (induction period method) is designed and manufactured according to GB / T8018 test method, It is suitable for determining the oxidation stability of gasoline under accelerated oxidation conditions.
Technical Parameters:
Operating temperature: 0 ℃ ~ 40 ℃
Bath temperature: 100 ± 0.1 ° C
Storage temperature: -20 ℃ ~ 50 ℃
Pressure range: 0 ~ 1.6MPa
Display: Color dot matrix LCD, 64 "
Display accuracy: 0.1 ℃
Display resolution: 640 * 480
Working voltage: AC220V, 50HZ
Oxygen intake pipe diameter: ф6
Machine power: 2700W
Weight: 50Kg
The main heating power: 1000W
The instrument uses computer control technology, automatic detection, calculation and control, the whole operation is simple
Control precision, high degree of automation and a series of advantages.
Test host uses the most advanced USB interface connected to the computer, easy to use.
Change the traditional water bath for the metal bath, no pollution, no water, more convenient to use.
Improve the design of the thermal insulation system, both to save energy and avoid the risk of burns to the operator.
The integrated design of the oxygen bomb test not only greatly shortens the oil and gas pressure test pipeline, makes the pressure detection more accurate, and because of the use of high sealing hand needle valve, can ensure good oxygen bomb during the test Sealed, so that the test results more reliable.
Oxygen Bracket design can facilitate the operator in its oxygen bomb on the demolition of the cover, but also facilitate the placement of oxygen bombs before and after the test.
During the test, various parameters of the system are displayed on the computer interface in real time. The pressure curves of the two oxygen bombs are also drawn in real time, and the test result is clear at a glance.
Automatic control of detection, real-time display of pressure data, automatic print pressure curve, the entire Chinese interface, metal bath heating, two bombs configuration, with notebook computers and printers
Heating tube power: 1600W, the actual heating power from the computer automatic control;
Metal bath temperature control point: 100 ℃ ± 1 ℃;
Oxygen bomb pressure transmitter measuring range: 0 ~ 1600kPa, accuracy ± 0.2 ‰
Label:

Service
Contact

Hotline

13304111506

Work Shift

from Monday to Friday

Office Number

0411-84794867

QR code